Top Menu

  • Search
  • ENQUIRY CART
  • Enquiry/Email
MENU
  • Main
  • Products
    • By Brands
      • All Brands
      • KYORITSU-Japan
      • SEEK Thermal-USA
      • Chauvin Arnouz (AEMC)-France
      • HIOKI-Japan
      • FLUKE-USA
      • DKK TOA-JAPAN
      • FLIR - SWEDEN
      • MADGETECH-USA
      • SEAWARD-UK
      • KESTREL-USA
      • GARRETT-USA
      • TESTO-Germany
      • TES-Taiwan
      • MEGGER-UK
      • LUTRON-Taiwan
      • DAVIS-USA
      • GARRETT-USA
      • GPI-Taiwan
      • Center-Taiwan
      • BW TECH-Canada
      • SEW-Taiwan
      • Extech-USA
      • Graphtec-Japan
      • NANOTRONIX-Korea
      • MITCORP-USA
      • DR FLU - USA
      • X VIDEOSCOPES - USA
      • FOTRIC - USA
      • MSR - SWITZERLAND
  • About KKInstruments
    • About KKInstruments
    • Our Customers
    • Proof of Purchases
    • Shop locations
  • Contact KKI
    • Enquiry/Contact us
    • Payment Methods
    • Forms
    • Shop locations
    • Support
      • Ways to buy
      • Warranty Period
      • Enquiry Form
      • Help
  • Shop locations
  • Enquiry Basket
  • NEW Body Thermometers
You are here: Home Products Airflow meters KIM110B - KKI AF560 AIRFLOW CAPTURE HOOD/BALOMETER W COLOR DIGITAL CLEAR DISPLAY (BUILT-IN TEMP/AIRFLOW/RH)

Client Testimonials

Well done, KKInstruments!
We managed to monitor our production process effectively with your expertise! Will buy again!

Ahmad Azizul
IR Chief Engineerof Exxon Mobil

Our company prefer to source measuring instruments only from KKInstruments because they have a wider range of products and show a high standard of expertise. We are very happy with the deal. No regrets for choosing KKInstruments for the past 8 years,

Joanna Low
Senior Engineerof Shell Malaysia Bhd

My Enquiry Cart

Your cart is empty

You have 0 items in your cart.

Show cart

Brands

Product Categories

  • 2G/3G/4G Wifi/Bluetooth (Cellular Activity Monitor)
  • About KKInstruments
  • Airflow Capture Hoods
  • Airflow meters
    • Airflow hoods
    • Airflow transmitters (Fixed types 2)
    • Hot-wire anemometers
    • Pitot tubes anemometers
    • Vane anemometers
    • Wind-cups anemometers
  • Airflow transmitters (Fixed types)
  • All New Additions
  • Appliance Checker
  • Barometers
  • Battery testers
  • Belt tension meters
  • Bench-top water testers
  • Brands
  • By Brands
  • Cable height meters
  • Cable Testers
  • Cable tracer/locator
  • Calibrators/Injectors
  • Calipers/Micrometers
  • Capacitance Decade Box
  • Capacitance Meter
  • Cases (Instrument Storage)
  • Clamp meters
  • Colour meters
  • Combustion / Flue Gas Analyzer
  • Concrete moisture meters
  • Cooking oil tester
  • Current Transformers (CT)
  • Datalogger/Recorders
  • Dataloggers WIFI
  • Decade Resistance Box
  • Densitometers
  • Depth gauge
  • Dew point meters
  • Digital Multimeter(DMM)
  • Discharge Sticks/Rods
  • Distance meters
  • Earth clamp meters/Earth resistance testers
  • ELCB/RCCB/RCD Testers
  • Emergency Mobility Lighting System
  • EMF/Radiation
  • Flaw detectors
  • Force gauge
  • Frequency Counter
  • Gas detectors/analysers/recorders
  • Gauss meters
  • Gloss meters
  • Handheld Weather Stations
  • Hardness tester
  • Heat Stress Meter
  • Height Gauge
  • High Voltage Testers
  • Hipot testers/Electrical Safety Analysers
  • Holiday detectors
  • Humidity meters (Hygrometers)
  • Hydrometers(Specific Gravity)
  • IAQ meters
  • Illuminance meters
  • Impedance testers
  • Inspection Cameras/Borescopes/VideoScopes/FiberScopes
  • Insulation (Resistance) testers
  • ION (Energy) meters
  • LCR meters
  • Light/Lux meters
  • Lightning detectors
  • Manifolds
  • Metal Scanners/Locators
  • Microscope
  • Milli-Ohm meter/Micro-Ohm meters
  • Moisture meters
  • Moisture Meters(Halogen)
  • Multifunction meters
  • Nuclear Radiation Monitor
  • Nutirents meters
  • Oil Analyser
  • Optical power meter
  • Oscilloscope/Scopemeters
  • OTDR(Time Domain Reflectometer)
  • Panel meters
  • Particle counters
  • Particle Discharge (PD) Tester
  • Peristaltic pumps
  • Phase sequence indicators
  • Phasing Sticks/Phasing Rods
  • Pitot tubes
  • Portable flow meter
  • Portable Medical Cooler/Fridge
  • Power analysers/Power meters
  • Power Supply
  • Pressure Instruments
  • Promotions
  • Protractors/Inclinometers
  • Radon analyser/Radon monitor
  • Rain gauge
  • Refractometer
  • RTU Telemetry Unit
  • SCBA (Breathing Apparatus)
  • Seed Counter
  • Seismograph
  • Smoke detectors
  • Socket testers
  • Soil moisture meters
  • Soil moisture/Soil ph meter
  • Solar Power Testers
  • Solar PV analysers
  • Sound level meters
  • Spectrophotometers(Colorimeter)
  • Static meters(Surface)
  • Stopwatch
  • Surface resistivity meters
  • Surface Roughness Testers/Surface Profile Gauge
  • Tachometer/Stroboscopes
  • Tension meters
  • Test Box/Trainers
  • Test hammers
  • Test-leads
  • Thermal imagers/thermal camera (Infrared camera)
  • Thermocouples wires
  • Thermometers
  • Thickness gauges(ultra-sonic)
  • Torque meters
  • Transformer Turns Ratiometers
  • Transmitters
  • U-Value meters
  • Ultra-Violet meters (UV meters)
  • Ultrasonic Flowmeters
  • Vibration meters
  • Voltage detectors
  • Volume Pricing Items
  • Wall scanners
  • Water
  • Water Activity Meters
  • Waveforn generator
  • Weather stations
  • Webster hardness testers
  • Welding pressure gauges
  • Whiteness meter
  • Wind meters
  • Windows tint meters (Reflectance meters)
  • Wireless Solution
  • Wood moisture meters

Contact KKI

For customer service:
Email: Sales@meter.com.my

Contact offices:

Johor Office (New 5000 square feet HQ): 
Suite 2, 9TH Floor (#9-02), Wisma Singlong, 9 Jln Zabedah, B.P 83000,Johor,MY.
Tel(Johor) : +607- 4352540, 4355540
Fax: +607-4347299

KL/PJ Office: 
207, Lvl 2, Blok B3, Leisure Commerce Square, Jln PJS8/9, 46150 Petaling Jaya, Selangor, MY. (Kurnia Insurance Building, Sunway)
Tel(KL) : +603- 78762460, 78762463
Fax(KL): +603-78762454

Penang Office:
Tel(PG): +6016-5505065

Office Hour:
Mon-Fri 9am to 6pm
Sat 9am to 1pm
Closed on Public Holidays

1
1
Screenshot_3
Screenshot_1
Screenshot_2

Products

Featured item KIM110B - KKI AF560 AIRFLOW CAPTURE HOOD/BALOMETER W COLOR DIGITAL CLEAR DISPLAY (BUILT-IN TEMP/AIRFLOW/RH)

Category: Airflow meters

 

Introduction and unsealing

The wind gauge has a touch screen interface that works with lithium batteries and AC power. Using the PC software in each product, you can download up to 5,000 pieces of data for analysis and reporting.

The air volume meter is suitable for air volume and temperature measurement in ventilation places such as air-conditioning pipes. The air volume cover can be used for air inlet and exhaust air outlet of any type of pipe after replacing different size hoods.

 

Product technical indicators

Model

AF560 wind capture hood

Sampling air volume range

100-3500m3/h

Sampling air volume accuracy

±5% of reading

Sampling air volume resolution

1 m3/h

Temperature measurement range

0-60℃

Temperature measurement accuracy

±0.5℃

Temperature measurement resolution

0.1℃

Humidity measurement range

0-100%RH

Humidity measurement resolution

0.1%RH

DC power supply

Two 18650 lithium batteries, 3.7V

AC power supply

AC220V 50Hz to DC9V

Battery life

Continuous standby for 12 hours

Wind speed

Wind speed can be read directly by air volume

Data memory capacity

1∽5000 sets of measurement data, can be queried

Test direction setting

A direction, B direction

Screen display

Color 4.3 inch touch screen

Printing method

External wireless thermal printer

Startup mode

Screen click, handle button, wireless remote control

Data communication interface

USB

surroundings

Operating environment: 0∽40°C Relative humidity 10∽90%

Storage environment: -30 ° C ∽ 45 ° C Relative humidity 0 ∽ 90%

Weight (with battery and 570 sample cover)

4Kg

Getting started

This chapter describes the characteristics, connections and installation methods of the wind meter. include:

  • Description of the instrument
  • Power supply
  • use of touch screen
  • Use of thermal printer
  • Use of a tripod
  • External interface use
  • Instrument description

The wind gauge has many convenient measurement functions. The power switch is located above and behind the right side panel. There are handles on both sides of the instrument base for easy grip.

  • Front picture

The control box is located in the middle of the handle.

  • Power supply

The air volume meter can be directly powered by a battery or externally connected to an AC power source.

Caution:

Please do not use the power adapter that is not configured by the instrument manufacturer to supply power. If it is light, the instrument will not be able to supply power normally. If it is heavy, the internal components of the instrument will burn out.

Warning:

Do not use other types of replacements on this instrument, as this may cause fire, explosion, and other hazards.

  • Use of touch screen

The wind gauges are sampled on the touch screen and can be operated with your fingertips. But don't use sharp objects such as pens or pencils to avoid damaging the touch screen. Also do not apply heavy pressure on the touch screen and damage the touch screen.

  • Use of wireless thermal printers

   The appearance is exquisite, the structure is light, the paper structure is easy to install, the rechargeable battery works, and the serial port, the USB interface and the bluetooth interface are provided. With standby sleep, automatic wake-up, black mark positioning, paper out detection, low-voltage alarm and other functions.

Operating

  • The printer is an easy-to-paper structure. Use both fingers to pinch the sides of the paper compartment cover to open the paper compartment cover, place the thermal paper roll (smooth face down), expose the paper head, close the paper compartment cover, and complete the paper loading.
  • POWER is the power button. In the battery power mode, press to turn it on, then press it to turn it off. In the adapter power mode, the printer is always on, and this button is invalid.
  • FEED is the paper feed button. When the power is on, press and hold this button, the printer will take the paper, if release, the paper feed to stop.

Printer maintenance and troubleshooting

In order to ensure that the printer works properly, be careful not to disassemble the printer head at will, and do not make any changes to the printer. For users who do not use the printer housing, pay more attention to protecting the handpiece.

  • Do not turn the printer on if the printer is not in use for a long time.
  • If the printer does not work properly, turn off the printer.
  • When replacing the paper roll, please pay attention to whether there is paper dust on the machine head. If it is, please remove it gently. The thermal paper should pay attention to the front and back, the reverse side is not coated, and the writing cannot be printed.
  • When the printer is printing or feeding, it cannot be torn; and it is also not possible to reverse the paper.。
  • Keep the printer control board clean and free of dust.
  • When the thermal printer does not print clearly, use a clean cotton ball with a little alcohol to gently wipe off the dirt on the surface of the heating element of the print head.
  • The thermal printer should choose a better quality paper when selecting the paper, which not only improves the printing effect, but also reduces the wear on the thermal film.
  • Use of external devices

USB cable

This standard USB interface is used to transfer data from the instrument to the computer through the PC software. The PC software can be saved into a data file. The data file is in XML format and can be opened in the EXL table. Please keep the instrument turned on during the data export process, and do not perform other operations, otherwise the data will be garbled.

  • Use of a tripod

The air volume meter is placed on the 20*20 aluminum profiled cross, and the locking edge of the base is locked with a locking screw to prevent the moving wind gauge from falling and causing damage.

Instructions

The wind gauge uses a touch screen display control. You can use your fingers, but don't use sharp objects (pens, etc.), which are likely to damage the screen.

Turn on the instrument and press the power switch. The boot screen of the product model and manufacturer name is displayed on the screen. The system can select the Chinese and English interface on this interface, and click to enter the sampling interface. As shown below:

 

 

 

 

When the sampling interface appears, the instrument is ready for operation.

  • sampling interface

The sampling interface is the main interface. The interface parameters are only displayable and cannot be changed. The customer can set the parameters in the setting interface before using the instrument.

  • This interface displays the sampled air volume. There is a start button and a wind direction icon on the lower side of the air volume (the picture is not displayed here, please refer to the real object). Click the start to test the current air volume, and the average value of the air volume appears within the set time.
  • The parameters such as the sampling position on the right side cannot be changed on this interface. It is only displayed. Before the test, you need to set the parameters of the current test in the setting interface, such as Location, Ser No., size, printing and storage function on and off. The display of wind speed is obtained by air volume conversion. If the size of the tuyere is not the current test size, the number displayed by the wind speed is not acceptable. Temperature and humidity show the real-time value of the current environment.
  • Query interface
  • Click “-” and “+” to query the data of the previous and next ones. You can click the middle of the red word in the sampling query box to pop up the number box. You can directly query the data by entering it.

 

 


 

Parts list

No.

Name

No.

Remarks

1

Bag

3

Main unit and main components one cloth bag, tripod and bracket one cloth bag, pulley one cloth bag

2

Air gauge base

1

Internal collection core, control box, etc.

3

Battery charger

1

9V

4

570*570Alloy frame

1

Cloth cover frame, assembled at the time of shipment

5

830*830Alloy frame

1

Cover frame, need to be assembled when using, see instructions

6

570*570 Cloth cover

1

 

7

830*830 Cloth cover

1

 

8

Aluminum alloy support rod 1

4

 

9

lithium battery

2

18650 battery (set in the instrument)

10

Wireless printer and charger

1

Already paired with the controller, ready to use with power on

11

Thermal printing paper

2

57*30mm

12

USB data cable

1

Used when exporting data in the control box

13

U disk

1

Contains PC software, drivers, etc.

14

tripod

1

Use with the bracket

15

Tripod pulley

1

Use with a tripod

16

Air gauge bracket

1

Use with a tripod

17

Instruction manual

1

 

18

testing report

1

 

19

Certificate

1

 

 

 

 

 

 

 

 

Ask about this product
« previous next »

Bottom Menu

  • Sitemap
  • Contact us
  • Products
  • Enquiry Cart
  • About KKInstruments
  • Latest Updated Website: Meter.com.my